Electrical Installation Tester
Mainte nance
69
Maintenance
XW Warnings
To prevent possible electrical shock, fire, or personal injury:
• Be sure that the battery polarity is correct to prevent battery
leakage.
• Repair the Product before use if the battery leaks.
• Have an approved technician repair the Product.
• Use only specified replacement parts.
• Replace a blown fuse with exact replacement only for
continued protection against arc flash.
• Do not operate the Product with covers removed or the case
open. Hazardous voltage exposure is possible.
• Remove the input signals before you clean the Product.
Periodically wipe the case with a damp cloth and mild detergent. Do not use
abrasives or solvents.
Dirt or moisture in the terminals can affect readings.
To clean the terminals:
1. Turn the meter off and remove all test leads.
2. Shake out any dirt that may be in the terminals.
3. Moisten a clean cotton swab with alcohol and clean the inside of each
terminal.
Table 20 is a list of replaceable parts for the Tester.
Table 20. Replacement Parts
Description
Part Number
W Fuse, 11 A, 1000 V 10.3 x 25.4 mm for Fused Probe
803293
W Fuse, 3.15 A, 500 V 6.35 x 32 mm for 166X Tester
2030852
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com