Specifications
DSO8000E Series HandHeld Oscilloscope User Manual 81
Chapter 11 Specifications
11.1 Technical Specifications
All specifications herein mentioned apply to the DSO8000E series oscilloscopes. Before checking
an oscilloscope to see if it complies with these specifications, make sure it meets the following
conditions:
The oscilloscope must have been operating continuously for twenty minutes under the
specified operating temperature.
The Do Self Cal operation must be performed through the Utility menu if the operating
temperature changes by more than 5.
The oscilloscope must be within the factory calibration interval.
All specifications are guaranteed unless noted ‘typical’.
Oscilloscope Specifications
Horizontal
Sample Rate Range
1GS/s
Waveform Interpolation
(sin x)/x
Record Length
Maximum 2M samples per single-channel; maximum 1M
samples per dual-channel (4K, 40K, 512K, 1M optional)
TIME/DIV Range
DSO8072E
DSO8102E
DSO8152E
DSO8202E
4ns/div to 2Ks/div, in a 2, 4, 8
sequence
2ns/div to 2Ks/div, in a 2, 4,
8 sequence
Sample Rate and
Delay Time Accuracy
±50ppm over any ≥1ms time interval
Delta Time Measurement
Accuracy
(Full Bandwidth)
Single-shot, Normal mode
± (1 sample interval +100ppm × reading + 0.6ns)
>16 averages
± (1 sample interval + 100ppm × reading + 0.4ns)
Sample interval = s/div ÷ 200
Position Range
DSO8072E DSO8102E
4ns/div to 8ns/div
(-8div × s/div) to 20ms
20ns/div to 80μs/div
(-8div × s/div) to 40ms
200μs/div to 2Ks/div
(-8div × s/div) to 2Ks
DSO8152E DSO8202E
2ns/div to 10ns/div
(-4div × s/div) to 20ms
Loading ...
Loading ...
Loading ...