Rohde & Schwarz MXO4-BNDL Oscilloscope

User Manual - Page 37

For MXO4-BNDL.

PDF File Manual, 40 pages, Read Online | Download pdf file

MXO4-BNDL photo
Loading ...
Loading ...
Loading ...
Rohde & Schwarz R&S®MXO4 Series Oscilloscope 37
Input and output
input frequency range 10 MHz (±20 ppm)
sensitivity
≥ –10 dBm into 50 Ω,
≤ 10 dBm at 10 MHz
Reference output connector BNC
impedance 50 Ω (nom.)
output signal
10 MHz (specified in timebase accuracy),
8 dBm (nom.)
Security slot for standard Kensington style lock
VESA mount
VESA compatibility mounting interface,
100 mm × 100 mm pattern size
Right side
Ground jack connected to ground
General data
Display
type
13.3" LC TFT color display with capacitive
touchscreen
resolution 1920 × 1080 pixel (Full HD)
Temperature
Temperature loading operating temperature range 0 °C to +50 °C
storage temperature range –40 °C to +70 °C
in line with MIL-PRF-28800F, section 4.5.5.1.1.1
class 3 tailored to +45 °C for operation
Climatic loading
+25 °C/+45 °C at 95 % relative humidity cyclic,
in line with IEC 60068-2-30
Altitude
Operating up to 3000 m above sea level
Nonoperating up to 4600 m above sea level
Mechanical resistance
Vibration sinusoidal
5 Hz to 150 Hz, max. 1.8 g at 55 Hz;
0.5 g from 55 Hz to 150 Hz, in line with
EN 60068-2-6
10 Hz to 55 Hz, in line with MIL-PRF-28800F,
section 4.5.5.3.2 class 3
random
10 Hz to 300 Hz, acceleration 1.2 g (RMS),
in line with EN 60068-2-64
5 Hz to 500 Hz, acceleration 2.058 g (RMS),
in line with MIL-PRF-28800F, section 4.5.5.3.1
class 3
Shock
40 g shock spectrum, in line with MIL-STD-810G,
method no. 516.6, procedure I
30 g functional shock, halfsine, duration 11 ms,
in line with MIL-PRF-28800F, section 4.5.5.4.1
Electromagnetic compatibility (EMC)
RF emission
in line with CISPR 11/EN 55011 group 1 class A
(for a shielded test setup);
instrument complies with emission require-
ments stipulated by EN 55011, EN 61326-1 and
EN 61326-2-1 class A, making it suitable for use
in industrial environments
Immunity
in line with IEC/EN 61326-1 table 2, immunity test
requirements for industrial environment
6)
Certifications
VDE,
C
CSA
US
, KC
Calibration interval
1 year
Power supply
AC supply
100 V to 240 V ±10 % at
50 Hz to 60 Hz and 400 Hz ±5 %,
max. 2.3 A to 1.3 A, in line with MIL-PRF 28800F,
section 3.5
Power consumption max. 210 W
Safety
in line with IEC 61010-1, EN 61010-1,
CAN/CSA-C22.2 No. 61010-1, UL 61010-1
6)
Test criterion is displayed noise level within ±1 div for input sensitivity of 5 mV/div.
Loading ...
Loading ...
Loading ...