Vertical system: analog channels
Input channels 4 channels
Input impedance
50 Ω ± 1.5 %,
1 MΩ ± 1 % || 12 pF (meas.)
Analog bandwidth (–3 dB) at 50 Ω input impedance
R&S®MXO4 ≥ 200 MHz
R&S®MXO4 with -B243 option ≥ 350 MHz
R&S®MXO4 with -B245 option ≥ 500 MHz
R&S®MXO4 with -B2410 option ≥ 1 GHz
R&S®MXO4 with -B2415 option
≥ 1.5 GHz
1)
at 1 MΩ input impedance
R&S®MXO4 ≥ 200 MHz (meas.)
R&S®MXO4 with -B243 option ≥ 350 MHz (meas.)
R&S®MXO4 with -B245 option ≥ 500 MHz (meas.)
R&S®MXO4 with -B2410 option
≥ 700 MHz (meas.)
2)
R&S®MXO4 with -B2415 option
≥ 700 MHz (meas.)
2)
Bandwidth limits max. –1.5 dB, min. –4 dB
1 GHz, 500 MHz, 350 MHz, 200 MHz, 100 MHz,
50 MHz, 20 MHz (meas.)
Rise/fall time (calculated)
10 % to 90 % at 50 Ω
R&S®MXO4 < 1.75 ns
R&S®MXO4 with -B243 option < 1 ns
R&S®MXO4 with -B245 option < 700 ps
R&S®MXO4 with -B2410 option < 350 ps
R&S®MXO4 with -B2415 option < 234 ps
Vertical resolution
12 bit,
18 bit for high definition mode (without reduction
of the sampling rate)
Input sensitivity at 50 Ω
0.5 mV/div to 1 V/div,
entire analog bandwidth supported for all input
sensitivities
at 1 MΩ
0.5 mV/div to 10 V/div,
entire analog bandwidth supported for all input
sensitivities
DC gain accuracy offset and position set to 0 V, after self-alignment
input sensitivity > 5 mV/div
±1 % full scale
input sensitivity ≤ 5 mV/div to ≥ 1 mV/div ±1.5 % full scale
input sensitivity < 1 mV/div ±2.5 % full scale
Input coupling at 50 Ω DC
at 1 MΩ DC, AC
Maximum input voltage at 50 Ω 5 V (RMS), 30 V (V
p
)
at 1 MΩ
300 V (RMS), 400 V (V
p
), derates at 20 dB/decade
to 5 V (RMS) above 250 kHz
at 1 MΩ with R&S®RT-ZP11 passive probe
400 V (RMS), 1650 V (V
p
),
300 V (RMS) CAT II;
for derating and details, see R&S®RT-Zxx Standard
Probes data sheet (PD3607.3851.22)
Position range ±5 div
Offset range at 50 Ω input sensitivity
100 mV/div to 1 V/div ±20 V
0.5 mV/div to < 100 mV/div ±5 V
Offset range at 1 MΩ input sensitivity
800 mV/div to 10 V/div ±200 V
80 mV/div to < 800 mV/div ±50 V
0.5 mV/div to < 80 mV/div ±(5 V – input sensitivity × position)
Offset accuracy
±(0.35 % ×
|net offset| + 0.5 mV + 0.1 div × input sensitivity)
(net offset = offset – position × input sensitivity)
DC measurement accuracy
after adequate suppression of measurement
noise using high-resolution sampling mode or
waveform averaging or a combination of both
±(DC gain accuracy ×
|reading – net offset|
+ offset accuracy)
1)
1.5 GHz analog bandwidth in interleave mode with 5 Gsample/s real-time sampling rate.
2)
With R&S®RT-ZP11 passive probe.