GW Instek MDO-2104EX Oscilloscope

USER MANUAL - Page 231

For MDO-2104EX. Series: MDO-2000E

PDF File Manual, 377 pages, Read Online | Download pdf file

MDO-2104EX photo
Loading ...
Loading ...
Loading ...
ARBITRARY WAVE GENERATOR
229
Type of Sweep
4. Press Type to set the sweep to
linear or logarithmic.
Type
Linear, Log
Start and Stop
Frequency
5. Press the Start or Stop soft-keys to
set the start and stop frequency,
respectively.
Start/Stop
25MHz ~ 0.1Hz
Note
To configure a up sweeping, set the start frequency at a
lower value than the stop frequency. To configure a
down sweeping, set the start frequency at a higher
value than the stop frequency.
Center Frequency
& Span
Alternatively the center frequency and span can be
set instead of the start and stop frequencies.
6. Press More 1 of 2.
7. Press Span to set the frequency
span of the sweep.
8. Press Center to set the center
frequency for the configured span.
Span
Center
25Mhz ~ -25MHz
25MHz ~ 0.1Hz
Note
To configure a up sweeping, set the span with a
positive frequency. To configure a down sweeping, set
the span with a negative frequency.
Loading ...
Loading ...
Loading ...