5.Menu Interpretation
55
of the current marker. This feature quickly moves the signal to the
center of the screen.
If Normal is selected, the center frequency will be set to the
frequency of the current marker.
If Delta Marker is selected, the center frequency will be set to the
frequency at which the Delta Marker is located.
The function is invalid in Zero span mode.
5.15.2 [Mkr->CF Step]
Sets the center frequency step of the analyzer based on the
frequency of the current marker.
If Normal is selected, the center frequency step will be set to the
frequency of current marker.
If Delta Marker is selected, this function is invalid.
The function is invalid in Zero span mode.
5.15.3 [Mkr->Start]
Sets the start frequency of the analyzer based on the frequency of
the current marker.
If Normal is selected, the start frequency will be set to the
frequency of the current marker.
If Delta Marker is selected, this function is invalid.
The function is invalid in Zero span mode.
5.15.4 [Mkr->Stop]
Sets the stop frequency of the analyzer based on the frequency of
the current marker.
If Normal is selected, the stop frequency will be set to the
frequency of the current marker.
If Delta Marker is selected, this function is invalid.
The function is invalid in Zero span mode.
5.15.5 [Mkr->Ref Level]
Sets the reference level of the analyzer based on the amplitude of
the current marker.
If Normal is selected, the reference level will be set to the
amplitude of the current marker.