4.User Notice
25
P Peak
K Marker
X File
L Save/Recall
E Marker Fctn
C Marker To
N Preset
H Help
J Measure
U Measure Set
F1~F7 F1~F7
F9 GHz/dBm F10 MHz/dB F11 kHz/dBmV F12
GHz/mv
0~9 0~9
Backspace <-
Esc X
Enter enter
4.10Basic Measurement
Basic measurements include, input signal frequency and amplitude
display, marked by a frequency marker. Follow these four simple
steps below to implement input signal measurement.
Basic:
1. Setting center frequency
2. Setting span and resolution bandwidth
3. Activate marker
4. Setting amplitude
For example, to measure a 100MHz -20dBM signal, you must turn
on the spectrum analyzer and ensure it is warmed up for 30
minutes to ensure measurement accuracy.
1. Equipment connection:
Connect the output terminal of signal generator to the RF Input
50Ω terminal of spectrum analyzer. Set the parameters as follows:
Frequency 100 MHz
Amplitude -20 dBm
2. Setting parameters:
1) Press Preset to restore the analyzer to its factory-defined