5.Menu Interpretation
38
5.5.3 [EMI Filter On Off]
Turn on or off EMI measurement resolution bandwidth.
If this menu item is under disabled state, it indicates that the EMI option
hasn’t been enabled for this model. Please enable it by registering option.
Currently, when opening an EMI filter (-6 dB bandwidth), the resolution
bandwidth can only be 200 Hz, 9 kHz, 120 kHz, or 1 MHz. At this time,
the detection method can be selected as "quasi-peak".
5.6 Trace
As the sweep signal is displayed as a trace on the screen, you
can set parameters about the trace using this key. The analyzer
allows for up to five traces to be displayed at one time, and press
this key to check the menu for trace line settings and operations.
5.6.1 [Trace 1 2 3 4 5 ]
Select trace. The spectrum analyzer provides traces 1, 2, 3, 4 and
5, and the number and state menu item of the selected trace will be
underlined. In particular, the color of number corresponds to the
color of trace.
5.6.2 [State ]
Set the refresh state type of spectral traces.
5.6.2.1 [Clear & Write]
Refresh the current spectrum curve by collecting real-time scanned
data at each point of the trace to display the analyzer trace.
5.6.2.2 [Max Hold]
Maintains the maximum for each point of the trace. It continuously
receive scan data and select positive peak value detect mode.
5.6.2.3 [Min Hold]
Maintains the minimum for each point of the trace. It continuously
receive scan data and select negative peak value detect mode.