120
Limit Bandwidth
Bandwidth limitation puts the input signal into a
selected bandwidth filter.
This function is useful for cutting out high
frequency noise to see a clear waveform shape.
The bandwidth filters available are dependent on
the bandwidth of the oscilloscope model.
1. Press the Channel key.
CH1
2. Press Bandwidth from the bottom
menu.
3. Choose a bandwidth* from the side menu
(depending on the bandwidth of the
oscilloscope).
Range
150MHz models: Full, 20MHz
250MHz models: Full, 20MHz,
100MHz
350MHz models: Full, 20MHz,
100MHz, 200MHz
500MHz models: Full, 20MHz,
100MHz, 200MHz, 350MHz
When the input impedance is set to 75Ω, the
bandwidth is limited to a maximum of 150MHz.