GW Instek MDO-2202EX Oscilloscope

User Manual - Page 370

For MDO-2202EX.

PDF File Manual, 377 pages, Read Online | Download pdf file

MDO-2202EX photo
Loading ...
Loading ...
Loading ...
MDO-2000E series User Manual
368
Probe Specifications
GTP-070B-4
Applicable to: MDO-2072EG/2072EX & MDO-2074EG/2074EX
Position x10
Attenuation Ratio
10:1
Bandwidth
DC to 70MHz
Input Resistance
10Mwhen used with oscilloscopes
with 1M input
Input Capacitance
14.5pF to 17.5pF
Compensation Range
10pF to 35pF
Max. Input Voltage
600V DC + ACpk
Position x1
Attenuation Ratio
1:1
Bandwidth
DC to 10MHz
Input Resistance
1M (oscilloscope input resistance)
Input Capacitance
85pF to 115pF
Max. Input Voltage
200V DC + ACpk
Operating Cond.
Temperature
-10°C to 50°C
Relative Humidity
85%
GTP-100B-4
Applicable to: MDO-2102EG/2102EX & MDO-2104EG/2104EX
Position X10
Attenuation Ratio
10:1
Bandwidth
DC to 100MHz
Input Resistance
10Mwhen used with oscilloscopes
with 1M input.
Input Capacitance
14.5pF to 17.5pF
Compensation Range
5pF to 30pF
Max. Input Voltage
600V DC + ACpk
Position X1
Attenuation Ratio
1:1
Bandwidth
DC to 10MHz
Input Resistance
1M(oscilloscope input resistance)
Input Capacitance
85pF to 115pF
Max. Input Voltage
200V DC + ACpk
Operating Cond.
Temperature
10°C to 50°C
Relative Humidity
85%
Loading ...
Loading ...
Loading ...