GW Instek MDO-2202EX Oscilloscope

User Manual - Page 364

For MDO-2202EX.

PDF File Manual, 377 pages, Read Online | Download pdf file

MDO-2202EX photo
Loading ...
Loading ...
Loading ...
MDO-2000E series User Manual
362
Polarity
Normal & Invert
Maximum Input Voltage
300Vrms, CAT I
Offset Position Range
1mV/div ~ 20mV/div : ±0.5V
50mV/div ~ 200mV/div : ±5V
500mV/div ~ 2V/div : ±25V
5V/div ~ 10V/div : ±250V
Waveform Signal
Process
+, -, ×, ÷, FFT, User Defined Expression
FFT: Spectral magnitude. Set FFT
Vertical Scale to Linear RMS or dBV
RMS or dBm, and FFT Window to
Rectangular, Hamming, Hanning or
Blackman.
Trigger
Source
CH1, CH2, CH3*, CH4*, Line, EXT**
*four channel models only.
**two channel models only.
Trigger Mode
Auto (supports Roll Mode for 100
ms/div and slower), Normal, Single
Sequence
Trigger Type
Edge, Pulse Width(Glitch), Video,
Pulse Runt, Rise & Fall(Slope),
Timeout, Alternate, Event-
Delay(1~65535 events), Time-
Delay(Duration, 4nS~10S), Bus
Holdoff range
4nS to 10S
Coupling
AC, DC, LF rej., Hf rej., Noise rej.
Sensitivity
1div
External
Trigger
Range
±15V
Sensitivity
DC ~ 100MHz Approx. 100mV
100MHz ~ 200MHz Approx. 150mV
Input Impedance
1MΩ±3%~16pF
Horizontal
Time base Range
1ns/div ~ 100s/div (1-2-5 increments)
ROLL: 100ms/div ~ 100s/div
Pre-trigger
10 div maximum
Post-trigger
2,000,000 div maximum.
Timebase Accuracy
±50 ppm over any 1 ms time interval
Real Time Sample Rate
1GSa/s max. (4ch models);
1GSa/s per channel (2ch models)
Record Length
Max. 10Mpts
Acquisition Mode
Normal, Average, Peak Detect, Single
Peak Detection
2nS (typical)
Average
selectable from 2 to 256
X-Y Mode
X-Axis Input
Channel 1; Channel 3*
*four channel models only
Y-Axis Input
Channel 2; Channel 4*
*four channel models only
Loading ...