MDO-2000E series User Manual
168
I
2
C Bus Trigger Settings
The I
2
C bus trigger conditions can be set at any time after the bus
settings has been set to I
2
C.
Panel Operation
1. Set the Bus to I
2
C in the bus menu.
Page 124
2. Press the Trigger Menu key.
Menu
3. Press Type from the bottom menu.
4. Press Others from the side
menu and select Bus.
The Trigger on settings will be reflected on the
Trigger Configuration icon.
From left: Bus trigger, Trigger source
5. Press Trigger On and select the
triggering condition for the
selected bus.
Trigger On
Start, Repeat Start, Stop, Missing
Ack, Address, Data, Address/Data
Trigger On Data
If Data or Address/Data was configured for the
Trigger On setting, then the number of bytes,
data and addressing mode (I
2
C) can be
configured.
6. Press Data from the bottom menu.