SDS5000X Series Digital Oscilloscope User Manual
Set the coupling mode of C1 to DC50Ω, and vertical scale to 500 mV/div,
vertical offset to 0. Set the trigger level to 0.
In normal mode, 5 pulses can be obtained on the screen with the sample rate of 1.25
GSa/s at the maximum memory depth.
Set the trigger mode to "Single", the timebase to 50 ns/div. Turn on the
Sequence mode, and set the segments to maximum (53,969 in this example,
up to 100,000 depending on the number of samples at the current time base).
Wait patiently until the acquisition completes, then all the waveforms satisfying
the trigger conditions are displayed onto the screen.