SDS5000X Series Digital Oscilloscope User Manual
18.10 Amplitude Strategy
According to different types of input signals, users can choose the
corresponding amplitude calculation strategy which can make the
measurement of top value and bottom values more accurate.
Touch ConfigAmplitude Strategy in the measure dialog box to recall the
amplitude strategy dialog box:
A. Set the amplitude calculation
strategy. When set to auto, the
amplitude calculation strategy
will be selected automatically
according to the input signal to
ensure the accuracy of the
measured value
B. Set the top value calculation
strategy. When set to histogram,
the value greater than half of
peak value will be counted, and
the value with the maximum
probability will be identified as
the top value; when set to max,
the maximum value of waveform
will be identified as the top value
C. Set the base value calculation
strategy. When set to histogram,
the value less than half of peak
value will be counted, and the