SIGLENT Remote Manual
93 SDM3055/SDM3055A
2.11.2 TRIGger:DELay {<seconds>|MIN|MAX|DEF} TRIGger:DELay?
[{MIN|MAX|DEF}]
Sets the delay between the trigger signal and the first measurement.
Parameter Typical Return
0 to ~3600 seconds (~1 μs steps), default
1 s +1.00000000E-06
Return five DC voltage measurements. There is a 2 second delay before each
measurement.
CONF:VOLT:DC 10
SAMP:COUN 5
TRIG:DEL 2
READ?
Typical Response
+3.07761360E-03,-1.16041169E-03,+5.60585356E-06,+1.21460160E-04,+2.85898
531E-04
Due to internal quantization, the actual delay that you set may be slightly different
than your specified value. The increment is approximately 1 μs. Use the query to
determine the exact delay. For example, if you send TRIG:DEL 500 ms and then
TRIG:DEL? the actual delay may be +5.00000753E-01.
By default, TRIGger:DELay:AUTO is ON. The instrument automatically determines
the delay based on function, range and integration time (see Automatic Trigger
Delays). However, you may need to set a delay longer than the automatic delay for
long cables, high capacitance or high impedance signals.
If you specify a trigger delay with this command, that delay is used for all functions
(except CONTinuity and DIODe) and ranges. The CONTinuity and DIODe tests
ignore the trigger delay setting.
If you have configured the instrument for more than one measurement per trigger
(SAMPle:COUNt >1), the delay is inserted after the trigger and between consecutive
measurements.
The instrument selects automatic trigger delay after a Factory Reset (*RST) or an
Instrument Preset (SYSTem:PRESet).