Siglent SDM3055 5½ Digit Dual-Display Digital Multimeter

Manual - Page 173

For SDM3055.

PDF File Manual, 236 pages, Read Online | Download pdf file

SDM3055 photo
Loading ...
Loading ...
Loading ...
SIGLENT Remote Manual
33 SDM3055/SDM3055A
2.5.2 CALCulate:AVERage:ALL?
CALCulate:AVERage:AVERage?
CALCulate:AVERage:COUNt?
CALCulate:AVERage:MAXimum?
CALCulate:AVERage:MINimum?
CALCulate:AVERage:PTPeak?
CALCulate:AVERage:SDEViation?
The CALCulate:AVERage:ALL? query returns the arithmetic mean (average), standard
deviation, minimum value and maximum value of all measurements taken since the
statistics were last cleared. The count and peak-to-peak statistics are not returned by
CALCulate:AVERage:ALL?.
The other six queries listed above return individual values.
Parameter Typical Return
(none) (see below)
Return the statistics of 100 frequency measurements:
CONF:FREQ
SAMP:COUN 100
CALC:AVER:STAT ON
INIT
CALC:AVER:ALL?
Typical Response:
-4.10466677E-04,+3.13684184E-04,+1.75743178E-02,-6.74799085E-04
Statistics are cleared when the measurement function changes or when any of these
commands is executed:
CALCulate:AVERage:STATe ON
CALCulate:AVERage:CLEar
INITiate
MEASure:<function>?
READ?
*RST
SYSTem:PRESet
All values except the COUNt are returned in the form +1.23450000E+01. The count is
as a signed, positive integer: +129.
When dB or dBm scaling is used, the CALC:AVER:AVER and CALC:AVER:SDEV
queries return +9.91000000E+37 (not a number).