Siglent SDM3055 5½ Digit Dual-Display Digital Multimeter

Manual - Page 183

For SDM3055.

PDF File Manual, 236 pages, Read Online | Download pdf file

SDM3055 photo
Loading ...
Loading ...
Loading ...
SIGLENT Remote Manual
43 SDM3055/SDM3055A
2.6.9 CONFigure:CAPacitance[{<range>|AUTO|MIN|MAX|DEF}]
Sets all measurement parameters and trigger parameters to their default values for
capacitance measurement. Also specifies the stalls through the incoming parameters.
Parameter Typical Return
<range>{2nF|20nF|200nF|2uF |20uF| 200uF|10000uF}
default AUTO (autorange)
none
Configure capacitance measurement using the 2uv range.Make and read two
measurements:
CONF:CAP 2uF
SAMP:COUN 2
READ?
Typical Response +7.26141264E-10,+7.26109188E-10
You can let autoranging select the measurement range, or you can manually select a
fixed range. Autoranging conveniently selects the range for each measurement
based on the input signal. For fastest measurements, use manual ranging
(autoranging may require additional time for range selection).
If the input signal is greater than can be measured on the specified manual range, the
instrument displays the word Overload on front panel and returns "9.9E37" from the
remote interface.
Use READ? or INITiate to start the measurement.
Loading ...
Loading ...
Loading ...