SDM3055/SDM3055A 20
2.3.1 CALCulate:TRANsform:HISTogram:ALL?
CALCulate:TRANsform:HISTogram:DATA?
The ALL form of the query returns a comma-separated list of the lower and upper range
values, the number of measurements, and the bin data collected since the last time the
histogram data was cleared. The DATA form returns only the bin data.
Parameter Typical Return
(none) See Example.
See Example.
The bin data includes the following, in order:
The number of measurements less than the lower range value .
The number of measurements in the each of the bins, starting at the lower range
value bin
The number of measurements greater than the upper range value
Range values are real numbers returned in the form +1.00000000E+00. The number
of measurements and bin data are signed, positive integers returned in the form
+100.