Specifications
DS O5000 Series Digital Storage Oscilloscope User Manual
67
Output Voltage,
typical
About 5V pp into 1M Ω load
Frequency, typical 1kHz
Power Supply
Supply Voltage
100-120VAC
RMS
( ± 10%), 45Hz to 440Hz , CAT
120-240VAC
RMS
( ± 10%), 45Hz to 66Hz , CAT
Power Consumption <30W
Fuse 2A , T rating, 250V
Environment al
Temperature
Operating: 32 to 122 ( 0 to 50 )
Nonoperating: -40 to 159.8 ( -40 to +71 )
Cooling Method Convection
Humidity
+104 or below ( +40 or below): 90% relative humidity
106 to 122 ( +41 to 50 ): 60% relative humidity
Altitude Operating and Nonoperating 3 , 000m ( 10 , 000 feet)
Random Vibration
0.31 g
RMS
from 50Hz to 500Hz , 10
minutes on each axis
Nonoperating
2.46g
RMS
from 5Hz to 500Hz , 10
minutes on each axis
Mechanical Shock Operating 50g , 11ms , half sine
Mechanical
Size
Length 313mm
Height 142mm
Depth 108mm
Weight
exclusive of packing and
accessories
2.08Kg
Packing
Length 3 85 mm
Width 200 mm
Height 2 4 5mm
Gross Weight inclusive of all accessories 2.5 Kg
8.2
8.2
8.2
8.2 Accessories
Accessories
Accessories
Accessories
All the following accessories are available by contacting your local HANTEK distributor .
Standard
Standard
Standard
Standard Accessories
Accessories
Accessories
Accessories
Sketch
Sketch
Sketch
Sketch Description
Description
Description
Description
X1 , X10 two p assive p robes. The passive probes have a 6MHz
bandwidth ( rated 100Vrms CAT III) when the switch is in the X1
position, and a maximum bandwidth ( rated 300Vrms CAT II) when
the switch is in the X10 position. Each probe consists of all
necessary fittings.
Loading ...
Loading ...
Loading ...