Delay
Set the time delay between the trigger input and the start of the N-Cycle burst.
Burst Trigger Source
The burst trigger source can be internal, external or manual. The generator
will generate a burst output when a trigger signal is received and then wait for
the next trigger source.
1. Internal Trigger
Choose Source Internal, the generator outputs continuous burst waveform
when internal trigger is selected. Choose Trig Out as ―Up‖ or ―Down‖, the [Aux
In/Out] connector at the rear panel will output a trigger signal with specified
edge.
2. External Trigger
Choose Source External, the generator accepts the trigger signal inputted
from the [Aux In/Out] connector at the rear panel when external trigger is
selected. A burst will be generated once the connector gets a CMOS pulse
with specified polarity. To set the CMOS pulse polarity, choose Edge to select
―Up‖ or ―Down‖.
3. Manual Trigger
Choose Source Manual, a burst will be generated from the corresponding
channel when the Trigger softkey is pressed when manual trigger is selected.