Measurement range Cuff pressure 0-300 mmHg,
Systolic 50-250 mmHg,
Diastolic 30-200 mmHg,
Pulse 40-180 beats/minute
Display accuracy Systolic ± 3 mmHg,
Diastolic ± 3 mmHg,
Pulse ± 5 % of the value shown
Measurement inac-
curacy
Max. permissible standard deviation
according to clinical testing:
Systolic 8 mmHg /
Diastolic 8 mmHg
Memory 2 x 60 memory spaces
Dimensions L 74 mm x W 72 mm x H 24.5 mm
Weight Approximately 100 g (without bat-
teries)
Cuff size
135 to 195 mm
Permissible operating
conditions
+5°C to +40°C, 15-93% relative humi-
dity (non-condensing)
Permissible storage
conditions
-25°C to +70°C, ≤ 93% relative humi-
dity, 700-1050 hPa ambient pressure
Power supply 2 x 1.5V
AAA batteries
Battery life For approx. 300 measurements, de-
pending on levels of blood pressure
and pump pressure
Accessories Instructions for use, 2 x 1.5V AAA
batteries, storage box
Classification Internal supply, IPX0, no AP or APG,
continuous operation, application part
type BF
This device complies with European Standard EN60601-1-2
and is subject to particular precautions with regard to electro
-
magnetic compatibility. Please note that portable and mobile
HF communication systems may interfere with this unit. More
details can be requested from the stated Customer Services
address or found at the end of the instructions for use.
The device complies with the EU Medical Devices Directive
93/42/EEC, the German Medical Devices Act (Medizinpro-
duktgesetz) and the standards EN1060-1 (Non-invasive sphy-
gmomanometers – Part 1: General requirements), EN1060-3
(Non-invasive sphygmomanometers – Part 3: Supplemen
-
tary requirements for electro-mechanical blood pressure
measuring systems) and IEC80601-2-30 (Medical electrical
equipment – Part 2-30: Particular requirements for the basic
safety and essential performance of automated non-invasive
sphygmomanometers).