Loading ...
Loading ...
Loading ...
46
IMPORTANT INFORMATION REGARDING ELECTRO MAGNETIC COMPATIBILITY EMC
Electromagnetic immunity IEC60601-1-2
Immunity test IEC 60601 Test level Compliance level
Electromagnetic environment –guidance
Voltage dips, short
interruptions and voltage
variations on power supply
IEC 61000-4-11
0 % U
T
for 0.5 cycle
0 % U
T
for 1 cycle
70 % U
T
for 25 cycles
0 % U
T
for 5 s
0 % U
T
for 0.5 cycle
0 % U
T
for 1 cycle
70 % U
T
for 25 cycles
0 % U
T
for 5 s
Mains power quality should be that of a typical home,
commercial or hospital environment. If user requires
continued operation during power mains interruption
insure that batteries are installed and charged. Insure
that battery life exceeds longest anticipated power
outrages or provide an additional uninterruptible
power source.
Power frequency (50/ 60
Hz) magnetic field
IEC 61000-4-8
30 A/m 30 A/m
Magnetic fields from common appliances are not
expected to affect the device.
Note: U
T
is the A.C. mains voltage prior to application of the test level.
Loading ...
Loading ...
Loading ...