Loading ...
Loading ...
Loading ...
24
GET STARTED WITH YOUR THERAPY
NEVER APPLY THE PADS TO THESE BODY AREAS:
Do not place the pads on or near
the head, mouth or face.
Do not place the pads over your
neck because this could cause
severe muscle spasms resulting in
closure of the airway, difficulty
in breathing, or adverse effects on
heart rhythm or blood pressure.
Do not use near the heart.
Do not place the pads on both sides
of the chest cavity simultaneously
(lateral or front and back), or
across your chest because the
introduction of electrical current
may cause rhythm disturbances.
Loading ...
Loading ...
Loading ...