26
Power frequency
(50/ 60 Hz)
magnetic field IEC
61000-4-8
3 A/m
Not applicable
Not applicable
Note: U
T
is the a.c. mains voltage prior to application of the test level.
Immunity test
IEC60601-1-2
test level
IEC60601-1-
2 test level
Electromagnetic environment - guidance
Conducted
RF IEC
61000-4-6
Radiated RF
IEC 61000-4-
3
3 Vrms 150
kHz to 80 MHz
80% AM (2Hz)
3 Vrms 80 MHz
to 2.5 GHz
80% AM (2Hz)
3 Vrms
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of BP A one
Max Connect, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommend separation distance
3V
d = 1.2×p
1/2
80Mhz to 800 MHz
d = 2.3×p
1/2
MHz to 2.5 GHz
Where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters as
determined by an electromagnetic site survey
a
,
should be less than the compliance level in each
frequency range
b
. Interference may occur in the
vicinity of equipment marked with the following
symbol:
Note1: At 80 MHz and 800 MHz, the higher frequency range applies.