24
Dimensions :
136×112×71±1.0 mm
Weight :
510 g±5g (including batteries and cuff)
Cuff pressure display range:
0~299mmHg/0~39.9kPa
Electrical shock protection:
Internal power unit
Safety classifications:
Type B equipment
Mode of operation:
Continuous operation
Protection against ingress of
water:
IP22
Accessories:
M-size Cuff , 4 “AA” batteries, instruction manual ,warranty card
Please be noticed the power adapter is not supplied from the origin, users can buy the adapter in the market
which must comply to EN60601-1, EN60601-1-2
14. Manufacturer’s Declaration
BP A one Max Connect is intended for use in the electromagnetic environment specified below. The customer
or the user of the BP A One Max Connect should assure that it is used in such an environment.
Electromagnetic Emissions: (IEC60601-1-2)
Emission Test
Compliance
Electromagnetic Environment
RF emission CISPR 11
Group 1
BP A one Max Connect uses RF energy only for
internal functions. Therefore, this RF emission is
extremely weak and there is little chance of it
creating any kind of interference whatsoever with
nearby electronic equipment.
RF emissions CISPR 11
Class B
BP A one Max Connect is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low voltage power supply network that
Harmonic emissions IEC
61000-3-2
Not applicable
Voltage fluctuations/flicker IEC
Not applicable