Welch Allyn ProBP 2400 digital blood pressure device
The ination pressure selection (mmHg/kPa) can be done in 1x,
, or Manual Mode. The default setting is “AUTO”
When set to AUTO, the ProBP 2400 will automatically determine
the maximum ination pressure as it inates the cuff.
Set maximum ination pressure
1) Press and hold the M Button for 3 seconds until the column
with pressure values ashes.
2)
Press the Start/Stop Button to select the preferred pressure
value (after value ‘260’ has been reached the next selection
option will be “AUTO” at the bottom of the list)
3) Press the M Button to conrm the selected value.
Push and hold for 3
seconds...
3 sec.
3 sec.
If the maximum ination pressure selected (or the
maximum ination pressure as determined while in
AUTO mode) is not adequate to determine systolic
pressure, the device will reinate to a pressure 30
mmHg higher than the previous ination pressure and
attempt another cycle. This can be repeated increasing
the maximum ination by 30 mmHg each time. If a
maximum ination pressure of 280 mmHg is reached,
and the ProBP 2400 is unable to determine a blood
pressure, an error code will be displayed.
It is recommended at this point to switch to the manual
mode and determine blood pressure with a stethoscope
using traditional Korotkoff method. (See «Manual»
Mode)