Loading ...
Loading ...
Loading ...
22
Special Functions
MAP (Mean Arterial Pressure)
The device measures the true mean arterial pressure
(MAP) of the patient. Each measurement includes a single
MAP value. The MAP value will always be displayed
together with the systolic and diastolic blood pressure
value.
Irregular heartbeat detector in «1x» Mode
The device detects irregular heartbeat in «1x» Mode.
The irregular heartbeat symbol shows up if during a
blood pressure measurement the heart rate has varied
by more than 25%. In case of an irregular heartbeat the
blood pressure measurement might be affected. It is
recommended to repeat the measurement or to choose
the «Manual» Mode for verication.
The Mean Arterial Pressure, MAP is determined from
the maximum peak of the oscillometric envelope curve.
Loading ...
Loading ...
Loading ...