Welch Allyn ProBP 2400 digital blood pressure device
Stand by
The user can manually select up measurement interval times
of 15, 30, 45 or 60 seconds in
Mode. (Please refer to special
functions section page 21 “Setting measurement interval times”).
The 60 second wait period before the rst measurement is not
adjustable but may be bypassed by pressing the Start/Stop
Button a second time. This will start the rst reading immediately.
« » Mode (automatic three reading average)
Select « » Mode then press the «Start/Stop» Button to
perform automatic triple blood pressure measurements to
determine a three reading average.
The countdown time before the rst measurement is set at 60
seconds.
The interval times between measurements is user adjustable to
15, 30, 45 or 60 seconds. The average measurement reading is
displayed and saved after the measurements are complete.
15, 30, 45,
or 60 sec.
15, 30, 45,
or 60 sec.
1 2
3
60 sec.
1x 3x 1x 3x
1x 3x
When the cuff pressure reaches 20mmHg during the
deation cycle, the remaining pressure is vented and the
ProBP 2400 goes into the Stand by mode.
Push the «Start/Stop» at any time to start fast deation and
set the ProBP 2400 to Stand by.
Take note – Take note of the systolic and diastolic values in
the same manner as performed with sphygmomanometer
measurements.
Set to ‘Stand by’ – The device can be set to ‘‘Stand by” by
pressing the «Start/Stop» Button without turning off the
power. The device will automatically switch to ‘Stand by’ if
there is no operation for one minute.