«Manual» Mode
Select «Manual» Mode if auscultatory blood pressure
measurement is preferred above using the oscillometric
method. In «Manual» Mode, the device serves only as a
pressure gauge. No oscillometric measurements will be
taken. The user can hear the systolic and diastolic Korotkoff
sounds by means of a stethoscope placed over the Brachial
artery.
Reusable ProBP 2400
Blood Pressure Cuff
22-32cm
ADULT
Reusable ProBP 2400
Blood Pressure Cuff
22-32cm
ADULT
Taking measurements in 1x, Manual and Mode (cont.)
Reusable ProBP 2400
Blood Pressure Cuff
22-32cm
ADULT
Start ination – Press the Start/Stop Button to start ination
of the cuff.
When maximum ination pressure is reached, the ProBP
2400 will automatically begin linear deation at a rate of
3mmHg/sec.
Re-inate –Push and Hold the M Button during deation
to re-inate for as long as the button is held up to a max
of 299mmHg. Release the button to continue deation.
Exceeding 299mmHg will result in an immediate release of
cuff pressure and a 'HI' Error message.