Specifications
Specifications
Artisan Sensor
Artisan Signal Cable
Test conditions
Reference-NIST certified reference
Ambient Temperature 24C +/- 2C
Display has been aged for >1 hour with a Full White field (85 cd/m2)
Display is calibrated to D50 (x = 0.346, y = 0.359, Y = 85 cd/m2)
Ambient light during calibration and measurement <2 lux
Signal timing--1152x870@70Hz
Accuracy
@85 cd/m
2
xy: +/- 0.002; Y: +/- 2%
@5 cd/m
2
<2 DeltaE deviation
@.5 cd/m
2
<2 DeltaE deviation
Repeatability
Repeatability in this context is: same display, same sensor, and same environmental conditions (measured by the same standard)
@85 cd/m
2
xy: +/- 0.002; Y: +/- 1%
@5 cd/m
2
xy: +/- 0.006; Y: +/- 3%
@0.5 cd/m
2
xy: +/- 0.006; Y: +/- 3%
Environmental Conditions
Operating Temperature 10 to 35 degrees C
Relative Humidity 20% to 80% non-condensing
USB
v.1.0
Power Consumption
5 Vdc +/-5%
20mA
Environmental Conditions
Operating Temperature 10 to 35 degrees C
Relative Humidity 10% to 80% non-condensing
USB
v.1.0
Power Consumption
5 Vdc +/-5%
20mA