CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 49/197
4.5.4 Water supply via freshwater can (optional water container set)
The fresh water can, which is placed in an included magnetic can holder on the left chamber side, serves
to supply one or several chambers with freshwater for humidification.
Since the pump is self-priming, it is not required that the water outlet of the can should be placed higher
than the pump, so you can position the water can freely in height.
The freshwater can outlet must NOT be placed on or above the rear panel of the chamber or on the right
chamber side.
DANGER
Electrical hazard by water entering the chamber via the rear panel.
Deadly electric shock.
The magnetic support adjustable in height (option) may only be attached on the left side
of the chamber.
Do NOT place the freshwater can tap on or above the rear panel of the chamber
Figure 20: Two CBF / CBF-UL incubators with installed water supply via freshwater can (option)