Binder CB 170 CO₂ incubator with hot air sterilization, series CB – 170-liter capacity

User Manual - Page 35

For CB 170.

PDF File Manual, 197 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 35/197
4.3 Connecting the O
2
sensor (chamber with O
2
control)
The O
2
sensor is supplied with the chamber in a separate package.
Connect or remove the O
2
sensor only when the chamber is turned off.
Open the door of the inner chamber and plug the O
2
sensor (F) into the left connection socket located in
the upper part of the rear of the inner chamber. Pay attention to the correct positioning of the pins.
Figure 12: O
2
sensor Figure 13: Connecting the O
2
sensor
The O
2
sensor must be plugged in during a hot-air sterilization.
4.4 Gas connections
General information for safe handling of gas cylinders:
Store and use gas cylinders only in well ventilated areas.
Open the gas cylinder valve slowly to avoid pressure surges.
Secure gas cylinders during storage and use against falling (chaining).
Transport gas cylinders with a cylinder cart, do not carry, roll, or throw them.
Always close the valve even with apparently empty cylinders; screw on the cap when not in
use. Return gas cylinders with the valve closed.
Do not open gas cylinders by force. Mark them when damaged.
Protect gas cylinders against fire, e.g. do not store together with flammable liquids.
Observe relevant regulations for dealing with gas cylinders.
Secure the gas cylinders against falling and other mechanical damage.
WARNING
Risk of injury through sudden release of the stored pressure energy when the valve
safety is torn off.
Injuries.
Secure gas cylinders against falling (chaining).
Transport gas cylinders with a cylinder cart.
The valve of the gas cylinder always must be closed before screwing on or unscrewing the gas hose.
WARNING
Risk of injury through sudden release of the stored pressure energy when opening
the cylinder valve of a not connected cylinder.
Injuries.
Close the gas cylinder valve before connecting or removing the gas hose.
Loading ...
Loading ...