KBF / KBF-UL + KMF (E6) 04/2020 page 76/163
11.1.2 Alarm messages
Condition Alarm message
Start after
condition occurred
Zero-voltage relay
alarm output (option)
The current actual temperature value
is outside the tolerance range (chap.
11.4)
Temperature
range
after configurable
time
time as alarm start
The current actual humidity value is
outside the tolerance range (chap.
11.4)
Humidity range
after configurable
time
time as alarm start
Open chamber door Door open after 5 minutes ----
Power failure --- ---- immediately
Exceeded setpoint of the safety control-
ler class 3.1
Safety controller immediately ----
Exceeded maximum or minimum tem-
perature (option temperature safety
device class 3.3)
Temp. safety
device
immediately ----
Temperature sensor defective
e.g. “- - - - “ or
“<-<-< or “>->->”
immediately ----
Safety controller temperature sensor
defective
Safety controller
sensor
immediately ----
Alarm messages are displayed in the list of active alarms until acknowledging them. They are also shown
in the event list.
11.1.3 Messages concerning the humidity system
Condition and measures Message
Start after condition
occurred
The humidity module is defective. Contact BINDER
service
Humidity system immediately
The humidity module cannot fill up.
In case of freshwater supply via water pipe:
The water tap is closed, or the chamber is defective
(e.g. inlet valve of humidity module).
In case of freshwater supply via freshwater can (option,
chap. 19.9):
Water can is empty. Humidification is turned off. In case
of refrigerating operation, the interior is strongly dehu-
midified. When the water supply is functional again, the
humidity system restarts, or the chamber is defective.
Freshwater supply immediately
The humidity module cannot empty the condensate
tank. Wastewater tube obstructed. Check the length and
location of the wastewater tube. If appropriate contact
BINDER service.
Wastewater immediately
Purging is required soon. Turn off and on again humidity
control in the “Setpoints” menu (chap. 6.3) to start purg-
ing. After a successful purging the notification resets
automatically.
Humidity system
purging req.
after predefined time
(approx. 3-5 months
depending on use)
Maintenance of the humidity system is required. Contact
BINDER service.
Humidity module
service
after predefined time
(approx. 1 year)
Messages concerning the humidity system are shown in the event list.