Binder KBF 115 Constant climate chamber with wide humidity range, liters

User Manual - Page 74

For KBF 115.

PDF File Manual, 163 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
KBF / KBF-UL + KMF (E6) 04/2020 page 74/163
10.6.4 Setpoint entry
Select the field “Temperature” and enter the desired temperature setpoint.
KBF / KBF-UL setting range: -5 °C up to 70 °C, KMF setting range: -15 °C up to 100 °C.
Confirm entry with Confirm icon. The controller returns to the section view.
Select the field “Humidity and enter the desired humidity setpoint.
KBF / KBF-UL setting range: 0% r.h. up to 80% r.h., KMF setting range: 0% r.h. up to 100% r.h.
Confirm entry with Confirm icon. The controller returns to the section view.
Select the field “Fan” and enter the desired fan speed setpoint.
Setting range: 40% up to 100% fan speed.
Confirm entry with Confirm icon. The controller returns to the section view.
10.6.5 Special controller functions via operation lines
You can define the switching state of up to 16 operation lines (control contacts). They are used to activate
/ deactivate special controller functions.
Operation line “Humidity off” serves to turn off the humidity.
Operation line Idle mode” activates / deactivates the operating mode “Idle mode”.
The other operation lines are without function.
Select the desired program and section. You can set the operation lines in the “Functions on/off” field.
For details please refer to chap. 9.7.3.
Loading ...
Loading ...
Loading ...