True-rms Digital Multimeter
General Specifications
47
Electromagnetic Compatibility (EN 61326-1:2005) ..... In an RF field of 3 V/M, accuracy = specified accuracy +20 counts, except 600 μA dc
range total accuracy = specified accuracy +60 counts. Temperature not specified
Relative Humidity .......................................................... 0 % to 80 % (0 °C to 35 °C)
0 % to 70 % (35 °C to 50 °C)
Battery Type ................................................................... 3 AAA Alkaline batteries, NEDA 24A IEC LR03
Approved Batteries ....................................................... Duracell Procell MN2400 LR03
Duracell Plus MN2400 LR03
Varta Max Tech No. 4703
Varta Industrial Alcaline No. 4003 (min. operating temperature is -10 °C)
Eveready Energizer No. E92
Rayovac Alkaline AAA (U.S. Type)
Panasonic LR03XWA
Battery Life ..................................................................... 400 hrs typical without backlight (Alkaline)
Vibration ......................................................................... Per MIL-PRF-28800 for a Class 2 instrument
Shock .............................................................................. 1 Meter drop per IEC 61010 (3 Meter drop with holster)
Size (H x W x L) .............................................................. 4.57 cm x 10.0 cm x 21.33 cm (1.80 in x 3.95 in x 8.40 in)
Size with Holster ............................................................ 6.35 cm x 10.0 cm x 19.81 cm (2.50 in x 3.95 in x 7.80 in)
Weight ............................................................................ 567.8 g (1.25 lb)
Weight with Holster and Flex-Stand ............................ 769.8 g (1.70 lb)
Safety Compliance ........................................................ Complies with ANSI/ISA S82.01-2004, CAN/CSA C22.2 61010-1-04 to 600 V
Measurement Category IV. Licensed by TÜV to EN61010-1, Pollution degree 2
Certifications ................................................................. CSA, TÜV, P,
, GOST, ATEX, IECEx
IP Rating ......................................................................... 67 (Non-operating. Protected against dust and the effect of immersion up to 1 m for
30 min.)