1586A
Users Manual
3-18
Reset the Channel and Test Configuration
To reset the entire channel configuration and test setup:
Note
All channels are set to the default settings when reset and cannot be
restored unless the configuration was saved.
1. Push and hold
for 3 seconds.
2. Push to confirm to reset channel setup, or push to cancel.
Analog Channel Configuration (Ch001, Ch101 to Ch222)
After an analog input is connected to the Input Module and/or the front-panel connectors,
the channel must be configured for the type of input that is connected. Analog inputs are
channels Ch101 to Ch122 and Ch201 to Ch222. Use the procedure below and the tables
in this section to set up channels for specific input types.
Please read the considerations below to help with channel configuration:
For new channels that have not been previously set up, the user is prompted to
configure the channel when it is set to ON for the first time. Push the cancel softkey
() while in the setup sequence to: (1) accept the default settings for new channels
or (2) use the previous configuration for channels previously configured.
A channel can be configured with the Edit Channel softkey in the Channel Setup
menu.
Each channel has independent channel options such as Channel Delay, Rate of
Change, Display As reading type, and thermocouple open-circuit detect that can be
accessed with the Channel Options softkey in the Edit Channel menu. The channel
options listed vary based on the on the input type.
After a channel is configured, it is important to verify the channel with the Verify
Channel softkey on the Channel Setup menu. This ensures that the channel was
configured properly before data is collected from a scan. See “Verify a Channel” on
page 3-14.
To configure an analog channel:
1. Push
.
2. Set the channel to ON (see “Set Channels to ON or OFF” on page 3-12). When the
channel is set to ON, a channel configuration sequence shows on the display that
steps the user through an initial configuration sequence as shown in Table 3-5. If the
channel was previously configured, push to cancel the configuration sequence
to use the last channel configuration.
3. Use the information in the tables in this section to configure the channel. Push or
to highlight the channel setting then push to edit the setting.