1586A
Users Manual
7-6
Table 7-1. Error Messages (cont.)
Error Code Error Message Causes and Solutions
Error 109 Unknown ACK byte
Cause: A packet was received, but it is not an
acknowledgement.
Solution: No user action necessary. The Instrument will
restart the GX.
Error 110
Quality indicator too
low
Cause: GX quality low.
Solution: No user action necessary. The Instrument will
restart the GX.
Error 111 GX restart
Cause: GX not responding.
Solution: No user action necessary. Instrument will restart
the GX.
Error 200
A/D conversion
timeout
Cause: A/D conversion timeout.
Solution: Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 201
Error detecting line
frequency
Cause: The line frequency is out of range.
Solution: Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 202 A2 memory failed
Cause: Failed to access memory on A2 board, or checksum
failure detected in the memory.
Solution: Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 203
Slot 1 module
memory failed
Cause: Failed to access memory on module of slot 1, or
checksum failure detected in the memory.
Solution: Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 204
Slot 2 module
memory failed
Cause: Failed to access memory on module of slot 2, or
checksum failure detected in the memory.
Solution: Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).