FLUKE 1586A/1DS/C Super-DAQ Precision Temperature Scanner with DAQ-STAQ multiplexer

User Manual for the Fluke 1586A/1DS/C Super-DAQ Precision Temperature Scanner with DAQ-STAQ multiplexer - Page 132

For 1586A/1DS/C.

PDF File Manual, 135 pages, Read Online | Download pdf file

1586A/1DS/C photo
Loading ...
Loading ...
Loading ...
1586A
Users Manual
7-18
Table 7-1. Error Messages (cont.)
Error Code Error Message Causes and Solutions
Error 910 File locations full
Cause: The number of setup file reaches the limit (99).
Solution: Remove some setup files.
Error 1302 Bad parameter count
Cause: The number of parameter does not match the
command.
Solution: Check the parameter and try again.
Error 1305 Illegal parameter unit
Cause: Illegal parameter unit was specified in the command.
The valid parameter unit is UV, MV, V, KV, UA, MA, A, OHM,
KOHM, MOHM, GOHM, TOHM, HZ, KHZ, MHZ, GHZ, CEL,
FAR.
Solution: Check the parameter and try again.
Error 1313
Operation only
allowed for service
Cause: The service only command was received while
service mode is not activated.
Solution: Enter service mode and try the operation again.
Error 1315 Invalid device trigger
Cause: An invalid bus trigger *TRG was received.
Solution: Input Ctrl+T when input buffer is not empty.
Error 1317
Communication input
buffer overrun
Cause: Input buffer overrun for communication.
Solution: Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 1320 Invalid binary data
Cause: An invalid binary data parameter was found in the
command.
Solution: Check the parameter and try again.
Error 1323 Invalid decimal data
Cause: An invalid decimal data parameter was found in the
command.
Solution: Check the parameter and try again.
Error 1326
Invalid hexadecimal
data
Cause: An invalid hexadecimal data parameter was found in
the command.
Solution: Check the parameter and try again.
Error 1328 Invalid octal data
Cause: An invalid octal data parameter was found in the
command.
Solution: Check the parameter and try again.