Channel Configuration
This section contains instructions on how to configure the channels after the inputs are
connected to the Input Module.
About Channel Numbers
A channel number (Ch) is a numerical identification associated with a set of terminals on
the Input Module. The channel number of the input is determined by the slot number the
Input Module is in (1 or 2) followed by the number of the terminal the input is connected
to (1 to 22) as illustrated in Figure 3-5. The front-panel inputs are permanently assigned
to channel Ch001. Here are some examples on how to determine the channel number of
an input:
A voltage source is connected to input terminal 4 (04) in the Input Module and then
slid into slot 1. The channel is Ch104.
A thermistor source is connected to input terminal 8 (08) in the Input Module and then
slid into slot 2. The channel is Ch208.
A voltage source is connected to the front-panel terminals. The channel is Ch001.
Table 3-2 shows channel types and channel numbers.
Table 3-2. Channel Types and Numbers
Channel Type
Channel Numbers
and Range
Reference
Input Channel (front panel) Ch001
See “Analog Channel Configuration (Ch001,
Ch102 to Ch222)” on page 3-18.
Input Channel (rear panel, slot 1) Ch101 to Ch122
Input Channel (rear panel, slot 2) Ch201 to Ch222
Digital I/O Channel Ch401
See “Digital I/O Channel Configuration
(Ch401)” on page 3-24.
Totalizer Channel Ch402
See “Totalizer Channel Configuration
(Ch402)” on page 3-25.
Math Channel Ch501 to Ch520
See “Math Channel Configuration (Ch501 to
Ch520)” on page 3-26.