1586A
Users Manual
7-16
Table 7-1. Error Messages (cont.)
Error Code Error Message Causes and Solutions
Error 704
Error reading
calibration folder
Cause: Internal flash is damaged.
Solution: Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 705
Error reading
calibration history
Cause: Internal flash is damaged.
Solution: Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 707
Error reading
calibration constants
Cause: Internal flash is damaged.
Solution: Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 708
Error writing
calibration constants
Cause: Internal flash is damaged.
Solution: Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 709
Error restoring LAN
settings
Cause: Internal flash is damaged.
Solution: Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 710
Error writing LAN
settings
Cause: Internal flash is damaged.
Solution: Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 711
Error setting MAC
address
Cause: Internal flash is damaged.
Solution: Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 712
Error storing MAC
address
Cause: Internal flash is damaged.
Solution: Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 800 Serial framing error
Cause: RS232 framing error
Solution: Make sure that the PC and Instrument settings are
the same.
Error 821 LAN port out of range
Cause: The LAN port < 1024 or > 65535.
Solution: Check the port address and try again.
Error 822
Error opening LAN
port
Cause: Error while opening LAN port.
Solution: Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 824
LAN port receive
error
Cause: Error while reading LAN port.
Solution: Check the LAN connection.