Industrial ScopeMeter®
Test Tool Setup
11
Measurement Connections
The Test Tool has two 4 mm safety-shielded banana jack
inputs (red input A and blue input B) and a black safety
4 mm banana jack input (COM). See Figure 4.
hxv05.eps
Figure 4. Input Connections for Measurement
Input A
Use the input A (red) for all single input measurements.
Input B
For measurements on two different signals, use B input
(blue) together with input A (red).
COM
Use the COM input (black) as the single ground for low
frequency, continuity, Ohm (Ω), diode, and capacitance
measurements.
XW Warning
To prevent electrical shock or fire, use only
one COM ˜ (common) connection, or ensure
that all connections to COM ˜ are at the
same potential.
Measurement Probe Setup
The Test Tool supports measurement probes, for
example:
10:1 voltage probe
1 mV/°C temperature probe
10 mV/A current clamp
To set up a specific probe type:
1. Push to open the Scope and Meter menu.
2. Push 3 to open the INPUT A menu.
3. Use Y/Z to highlight SELECT.
4. Push to open the PROBE select menu.
5. Use
Y/Z to highlight the probe type.
6. Push to accept the probe type and close the
menu.