Ti32, TiR32, Ti29, TiR29, Ti27, TiR27
Users Manual
26
2. Press the softkey labeled Up or Down to move between the materials in
the list. The emissivity value for each material is shown on the screen as
the selection moves between the different materials
3. Press the softkey labeled Done to select the highlighted material.
To set the emissivity value directly:
1. Press the softkey labeled ε.
2. Press the softkey labeled Up or Down to increment or decrement,
respectively, the emissivity value displayed just above the softkey labels.
3. Press the softkey labeled Done to select the set value.
4. Wait for main menu to disappear or quickly squeeze and release the
trigger twice in order to return to live view.
Note
If you have the Display Information setting on your Imager set to
Display All, information on current emissivity settings can be see as
ε
= xx”.
Setting Reflected Background Temperature
(Reflected Temperature Compensation)
Compensation for reflected background temperature on the Imager is set in the
Background tab. Very hot or very cold objects may affect the temperature
measurement accuracy of the target or object of interest, especially when
surface emissivity is low. Adjustment of the reflected background temperature
setting may improve temperature measurement accuracy.
1. Press until Background appears over .
2. Press softkey labeled Background.
3. Use the softkey labeled Up or Down to adjust the reflected background
temperature.
4. Press Done when finished.
5. Wait for main menu to disappear or quickly squeeze and release the
trigger twice in order to return to live view.