772/773
Calibration Manual
16
mA DC Clamp Measure Adjustment Procedure
To adjust the Clamp Measure function, use the Calibrator to apply the necessary
Meter input and do the steps in Table 6.
Table 6. Clamp Measure Adjustment Procedure
Step Meter Display Calibrator Output Procedure
1 0.00 mA 0.00 mA Stop for 10 seconds, push
2 -20.00mA -20.00 mA Push
3 0.00mA 0.00 mA Stop for 10 seconds, push
4 20.00mA 20.00 mA Push
5 0.00 mA 0.00 mA Stop for 10 seconds, push
6 -100.00mA -100.00 mA Push
7 0.00mA 0.00 mA Stop for 10 seconds, push
8 10.00mA 100.00 mA Push
9 Save -- Push
mA DC Measure Adjustment Procedure
To adjust the mA Measure function, use the Calibrator to apply the necessary
Meter input and do the steps in Table 7.
Table 7. mA Measure Adjustment Procedure
Step Meter Display Calibrator Output Procedure
1 -20.00 mA -20.00 mA Push
2 0.00 mA 0.00 mA Push
3 20.00 mA 20.00 mA Push
4 Save -- Push
Volts DC Measure Adjustment Procedure
To adjust the Volt Measure function, use the Calibrator to apply the necessary
Meter input and do the steps in Table 8.
Table 8. Volt Measure Adjustment Procedure
Step Meter Display Calibrator Output Procedure
1 -30.00V -30.00 V Push
2 0.00V 0.00 V Push
3 30.00V 30.00 V Push
4 Save -- Push