FLUKE 773 CAL Milliamp Process Clamp Meter with calibration certificate

User Manual for the Fluke 773 CAL Milliamp Process Clamp Meter with calibration certificate - Page 13

For 773 CAL.

PDF File Manual, 21 pages, Read Online | Download pdf file

773 CAL photo
Loading ...
Loading ...
Loading ...
772/773
Calibration Manual
10
gep02.emf
Figure 2. Battery Replacement
Performance Tests
XWWarning
To prevent electrical shock, personal injury, or fire:
Repairs or Meter servicing must be done only by
qualified
personnel.
Do not do the verification tests or calibration adjustment in
this manual unless qualified.
The tests th
at follow verify the functions of the Meter. If the Meter fails the
verification tests, repair is necessary. For Meter servicing, see How to Contact
Fluke.
Required Equipment
Required equipment for the performance tests is in Table 2. If the recommended
models are not available, equipment with equivalent specifications can be used.
Table 2. Required Equipment
Equipment Minimum Required Characteristics Recommended Model
Calibrator DC milliamps:
0-24.00 mA = ±0.073 %
24.0-100.0 mA = ±0.375 %
DC Volts:
0-30.00 V = ±0.267 %
Fluke 55xxA
DMM DC Current:
0-24.00 mA = ±0.375 %
DC Volts:
0-10 V = ±0.1 %
Fluke 88xxA
Lab Supply 6 Vdc ±-0.5 V -
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com