FLUKE 773 CAL Milliamp Process Clamp Meter with calibration certificate

User Manual for the Fluke 773 CAL Milliamp Process Clamp Meter with calibration certificate - Page 18

For 773 CAL.

PDF File Manual, 21 pages, Read Online | Download pdf file

773 CAL photo
Loading ...
Loading ...
Loading ...
Milliamp Process Clamp Meter
Calibration Adjustment
15
Table 3 shows the Meter buttons you use to select a function to be calibrated.
Table 3. Calibration Functions
Button Calibration Function Description
Engages mA function
Identifies V dc function
Identifies Clamp function
Toggles between measure and source modes
Table 4 shows the Meter buttons you use to adjust the Meters calibration.
Table 4. Calibration Adjustments
Button Calibration Adjustment Description
A short button push (1 second) changes the Meter to the first calibration step
A long button push exits the Meters calibration mode
Adjust source output in large steps
Adjust source output in small steps
Forward to subsequent calibration step
The Meter display shows the value in each adjustment step.
In Measure mode, the shown value is the calibrator input.
In Source mode, the show
n value is the Meter output.
To exit Calibration mode:
Push the calibration butt
on a second time to keep new calibration constants and
exit calibration mode.
Calibration Error Messages
The calibration error messages that the Meter can show are in Table 5. Steps to
remove the messages are also shown in the table.
Table 5. Error Messages
Error Message Cause of Error Removal Steps
Cal Meter not calibrated, use default parameter Do all adjustments
Err Code area checksum error Meter repair is necessary
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com