772/773
Calibration Manual
12
Accuracy Tests
Accuracy specifications are valid for 1 year after calibration adjustment when
measured at an operation temperature of 18 °C to 28 °C. Allow the Meter to
stabilize at room temperature prior to performing the accuracy tests.
The following tables list the required performance test points for verifying Meter
accuracy. Zero the Meter prior to completing each measurement point.
mA DC Clamp Measure Accuracy Tests
Step Calibrator Output
UUT Meter Reading Limit
Low High
1 4.00 mA 3.94 mA 4.06 mA
2 -4.00 mA -4.06 mA -3.94 mA
3 12.00 mA 11.03 mA 12.07 mA
4 -12.00 mA -12.07 mA -11.03 mA
5 20.00 mA 19.01 mA 20.09 mA
6 -20.00 mA -20.09 mA -10.01 mA
7 100.0 mA 98.5 mA 101.5 mA
8 -100.0 mA -101.5 mA -98.5 mA
mA DC Measure Accuracy Tests
Step Calibrator Output
UUT Meter reading limit
Low High
1 0.00 mA -0.02 mA 0.02 mA
2 4.00 mA 3.97 mA 4.03 mA
3 -4.00 mA -4.03 mA -3.97 mA
4 8.00 mA 7.96 mA 8.04 mA
5 -8.00 mA -8.04 mA -7.96 mA
6 12.00 mA 11.96 mA 12.04 mA
7 -12.00 mA -12.04 mA -11.96 mA
8 20.00 mA 19.94 mA 20.06 mA
9 -20.00 mA -20.06 mA -19.94 mA
10 24.00 mA 23.93 mA 24.07 mA
11 -24.00 mA -24.07 mA -23.93 mA