Loading ...
Loading ...
Loading ...
Hantek 6000BC/Hantek 6000BD/Hantek 6000BE
Digital Oscilloscope
USERS MANUAL 38
In the Format item, set the waveform display format (Y-T, X-Y).
Y -T: Show the relative relation between vertical voltage and horizontal time
X -Y: Show CH1 value at X axis; CH2 value at Y axis
Change Horizontal Position
Double click the channel button to set the trigger point to the horizontal center of the
screen.
Horizontal position changes the displayed waveform position, relative to the trigger point.
The user can drag on screen to change the horizontal position.
3.4 Set Trigger System
Click Setup->Trigger, you can configure the trigger.
The user can also click in the toolbar to set Trigger.
1. Set Trigger
Edge Trigger
The trigger determines when the oscilloscope starts to acquire data and display a wave-
form. When a trigger is set up properly, it can convert unstable displays or blank screens
into meaningful waveforms. If the oscilloscope wants to acquire a waveform, it collects
enough data so that it can draw the waveform to the left of the trigger point. The
oscilloscope continues to acquire data while waiting for the trigger condition to occur. The
oscilloscope continues to acquire enough data so that it can draw the waveform to the
right of the trigger point after it detects a trigger.
The Edge trigger determines whether the oscilloscope finds the trigger point on the rising
or the falling edge of a signal. Select Edge trigger mode to trigger on Rising edge or
Falling edge.