Table 7-1. Error Messages (cont.)
Error Code Error Message Causes and Solutions
Error 407
Reference channel
not enabled
Cause: TRIGger:AUTO:REF or TRIGger:AUTO:SREF
command is attempting to configure an OFF channel as
reference.
Solution: Configure the channel which you want to use as
reference.
Error 408
Channel used as
reference
Cause: A channel cannot be set to OFF while it is used as
reference channel for automated test.
Solution: Change the reference channel setting or
reconfigure the automated test.
Error 409
Reference channel
not set
Cause: Automated test is initiated, but reference channel has
not been assigned.
Solution: Assign a channel as reference channel.
Error 410
Alarm channel not
set
Cause: A scan is started but the alarm trigger does not have
an alarm channel assigned.
Solution: Assign a channel as alarm trigger channel.
Error 411 Pause ignored
Cause: Invalid pause operation while scan has not been
initiated.
Solution: Pause is only allowed when scan is initiated.
Error 412 Scan list is empty
Cause: No channel is ON while scan is being initiated.
Solution: Configure the channels before the scan is started.
Error 414
Current function not
available in module
Cause: Chx21 or Chx22 is enabled in the configuration, but
the installed Input Modules do not include these channel IDs
or the wrong setup file or wrong Input Module is installed.
Solution: Check to make sure the Input Modules are
installed correctly and registered. See Chapter 2.