Error Messages and Troubleshooting
Error Messages 7
7-5
Table 7-1. Error Messages (cont.)
Error Code Error Message Causes and Solutions
Error -102 Syntax error
Cause: Invalid syntax was found in the command string.
Solution: Check the parameter value and try again.
Error 0 No error Cause: Error queue is empty. No action necessary.
Error 100
Acknowledgement
queue full
Cause: GX acknowledgement queue overflow.
Note
GX is the abbreviation for Guard Crossing - which is the
communication link between the inguard and outguard.
Solution: Instrument will restart the GX.
Error 101
Inguard not
responding (recv)
Cause: Inguard not responding (receive).
Solution: Instrument will restart the GX.
Error 102
Lost sync with
inguard
Cause: Lost sync with inguard.
Solution: Instrument will restart the GX.
Error 103
Inguard not
responding (send)
Cause: Inguard not responding (send).
Solution: Instrument will restart the GX.
Error 104
Oversize packet
received
Cause: Oversize packet received.
Solution: The packet will be ignored automatically.
Error 105 GX CRC error
Cause: CRC error detected.
Solution: No user action necessary. The packet will be
ignored automatically.
Error 106 Wrong ACK number
Cause: An ACK packet was received with unexpected ACK
number.
Solution: No user action necessary. The Instrument will
restart the GX.
Error 107
Info packet received;
link not active
Cause: An INFO packet was received, but the
communication has not been setup.
Solution: No user action necessary. The packet will be
ignored automatically.
Error 108 Unknown control byte
Cause: A packet was received with unknown control byte.
Solution: No user action necessary. Instrument will restart
the GX.