11 Maintenance
The instrument has been designed with the utmost care. Ease of operation
and simplicity of maintenance have been a central theme in the product
development. Therefore, with proper care the instrument should require
very little maintenance. Avoid operating the instrument in an oily, wet,
dirty, or dusty environment.
Ice will build up over a period of time if the unit is operated at sub-zero
temperatures, allowing water to pool in the well at temperatures above
0.0°C. The user needs to drain the water build up after every use. Two
methods are available for draining the water: 1) Remove the insert from
the well and tip the unit upside down draining all water. 2) Take the unit
above 100°C for a period of one hour causing the water to evaporate.
If the outside of the instrument becomes soiled, it may be wiped clean
with a damp cloth and mild detergent. Do not use harsh chemicals on the
surface which may damage the paint.
Be sure that the well of the calibrator is kept clean and clear of any for-
eign matter. DO NOT use fluids to clean out the well.
The instrument should be handled with care. Avoid knocking or dropping
the instrument.
If a hazardous material is spilt on or inside the instrument, the user is re-
sponsible for taking the appropriate decontamination steps as outlined by
the national safety council with respect to the material.
If the mains supply cord becomes damaged, replace it with a cord of the
appropriate gauge wire for the current of the instrument. If there are any
questions, call an Authorized Service Center for more information.
Before using any cleaning or decontamination method except those rec-
ommended by Hart, users should check with an Authorized Service Cen-
ter to be sure that the proposed method will not damage the equipment.
If the instrument is used in a manner not in accordance with the equip
-
ment design, the operation of the calibrator may be impaired or safety
hazards may arise.